The annual self-audit is best run by someone senior enough to see the whole system but not the person who runs it day to day: a quality manager, a deputy, or a senior embryologist from another part of the laboratory. The person who maintains the vessels will unconsciously score their own workarounds as passes. Full independence is not required for a self-audit; its purpose is to find problems cheaply before an external inspection or a failure finds them expensively. What should be independent is the evidence: the auditor checks the fill log plot, triggers the alarm test, and pulls the spot-check goblets personally rather than accepting verbal assurance. Our self-audit guide is written to be handed to whoever draws the task.