The monitor belongs in the room where the nitrogen lives, positioned where people actually breathe, and guidance used across UK institutions sets the first alarm at 19.5 percent oxygen, against 20.9 percent in normal air. Two placement details matter. First, freshly boiled-off nitrogen is cold and denser than room air, so it pools low before mixing; a sensor at both working height and low level covers the real behaviour of a spill rather than the textbook behaviour of warm nitrogen. Second, the display or repeater should be visible from OUTSIDE the door, because the whole point is to warn someone before they walk in; University of Bath’s cryogenic guidance, built on the British Compressed Gases Association codes, treats the ventilation and monitoring assessment as the first control for any storage area. Size the risk honestly: a single small dewar in a large ventilated laboratory may not need one, but a store room holding several liquid nitrogen tanks, or any room where filling happens, almost always does.