Enough to prove, years later, that every vessel held temperature every day, and enough to buy time when one does not. The working minimum is an independent temperature probe in each vessel logging continuously to a system that archives the data for the life of the stored material, not the life of the equipment, plus level monitoring and alarms that reach a person out of hours. Independent matters: the probe that raises the alarm should not share a failure mode with the vessel it watches, and alarm response should be a written procedure with names on it, not an assumption.

The reason the alarm chain deserves this attention is that cryogenic failure is slow when you understand it and catastrophic when you do not. A vacuum jacketed vessel that loses its vacuum gives a bank hours to act, and our analysis of the timeline from vacuum failure to sample loss shows where those hours go. Vapour phase vessels also have a temperature structure worth understanding before you place probes, covered in our guide to liquid nitrogen storage temperature. Monitoring then needs auditing like everything else: our annual cryo storage self audit includes the alarm tests most banks skip, and scheduled cryogenic equipment servicing keeps the vessels themselves inside their performance envelope. The wider storage architecture sits in our article on cord blood and stem cell bank storage.