First, check these common issues:
Problem: Temperature above -190°C
Common causes:
- Vacuum seal failure – Most common
- Symptom: Rapid temperature rise
- Fix: Recertification/replacement needed
- Cost: £50-150 for testing
- High evaporation rate – Partial vacuum loss
- Symptom: Increased refill frequency
- Fix: Vacuum re-certification
- Cost: £100-200
- Refill interval exceeded – Normal operation
- Symptom: Temperature drops as LN₂ evaporates
- Fix: Schedule refill sooner
- Cost: £20-35 per refill
What to do:
- Check when last refilled (if >8 weeks, refill immediately)
- Verify refill port is functioning (try adding small amount)
- Contact supplier for professional inspection
- Use dry shipper as emergency backup if available
Sample safety:
- Samples remain viable down to approximately -150°C
- Even failed dewars maintain samples for 30+ days