Choose the depth your validated method is written around, then keep it fixed. A 10 µm chamber is commonly used in human-focused laboratories and a 20 µm chamber is commonly used in veterinary and research settings, and both depths are available in the SpermScope® range.
Depth matters because a chamber converts a count into a concentration through the volume of liquid sitting above the grid. For a given grid, doubling the depth doubles that volume, so the same raw count represents half the concentration. In the method described in our validation study article, 10 microlitres are loaded under a circular coverslip giving an estimated depth of around 10 micrometres, all 100 squares are counted and the total is divided by ten to give millions per millilitre. Apply that divisor to a 20 µm chamber and the result is wrong by a factor of two, a mistake that looks completely plausible on the worksheet.
That is why depth is a procedure decision, not a purchasing preference. Write the depth into the SOP, record it with each result, and confirm the conversion factor with the manufacturer’s instructions for the chamber in use. Laboratories running both human and research work should store the two depths separately and label them clearly, so the wrong chamber is never picked up on a busy treatment or freezing day. Both depths carry an individual serial number, which makes it straightforward to show which chamber was used for which assessment.