Both, for different purposes.

Nitrogen boils at minus 195.8 degrees Celsius at atmospheric pressure, so that is the precise figure and it is what a sample immersed in the liquid sits at. Minus 196 is the rounded version and it is what most documentation and most protocols use. Either is defensible and there is no point arguing about the difference.

The error worth avoiding is different. Neither number describes a vapour phase store. In vapour phase the samples sit above the liquid in cold gas, with a vertical temperature gradient up the vessel, so a single figure does not describe the store at all.

Quote the liquid temperature when you mean the liquid. When you mean vapour phase, quote the warmest position a sample can occupy, because that is the number the storage has to satisfy.