The hold time of a cryogenic tank – the number of days it maintains its liquid nitrogen charge without replenishment – depends on the vessel’s vacuum insulation quality, its static evaporation rate, its capacity relative to the fill level, ambient temperature, and how frequently the vessel is accessed. A large-format storage vessel in a climate-controlled laboratory with a low static evaporation rate may hold its charge for several weeks to months. A dry shipper used for transport will typically hold its charge for 5 to 14 days under normal conditions.
What is static evaporation rate and why does it matter?
Static evaporation rate is the volume of liquid nitrogen a vessel loses per day without any access or movement. It is expressed in litres per day and is one of the primary performance specifications for a cryogenic storage tank. A lower static evaporation rate means longer hold times between top-ups, lower liquid nitrogen consumption costs, and greater operational resilience during supply disruption. When comparing vessels, static evaporation rate is as important as canister capacity.
Does opening a cryogenic tank affect its hold time?
Yes. Every time a cryogenic tank is opened and a canister is raised into the neck, warm ambient air enters the vessel and accelerates liquid nitrogen evaporation. High-access laboratories – those retrieving or adding samples multiple times per day – will experience measurably higher evaporation rates than the published static rate. This should be factored into top-up scheduling.
What happens if a cryogenic tank runs dry?
If liquid nitrogen is not replenished before the vessel runs dry, the temperature within the vessel will rise above -130 degrees C – the critical threshold below which biological processes are suspended. Above this threshold, cellular deterioration resumes. If temperatures rise to -80 degrees C or above for any sustained period, sample viability is likely to be compromised. This is why continuous temperature monitoring and nitrogen level alarms are standard in licensed IVF units.
The published hold time of a cryogenic tank is a static measurement taken under ideal laboratory conditions. In a busy IVF unit with daily access, the actual hold time will be shorter. Calculate your top-up schedule from your measured evaporation rate, not the product datasheet.
Cryolab’s CryoNest storage vessels are available with detailed evaporation rate specifications. Browse at cryolab.co.uk/product-category/storage-vessels/.