Run the worst case sum before the first vessel arrives. Take the total liquid volume the room could hold, multiply by roughly 680 to get the gas volume a complete loss would release, and compare it with the room volume: if a full spill could push the oxygen concentration below 19.5 percent, the room needs forced ventilation, an oxygen depletion alarm, or a different location. A 30 litre dewar failing in a 4 by 3 metre store room is a very different event from the same dewar in an open laboratory. Then check the slow case as well, because vessels boil off continuously, and a tightly sealed room with poor air changes can drift oxygen-deficient over days with nothing having gone wrong. Basements, cellars, lift lobbies and windowless cupboards are the classic mistakes; Trinity College Dublin’s cryogen safety guidance covers the assessment logic well. Vessel choice is part of the answer too: a storage tank with a healthy vacuum and the right neck design releases a fraction of the boil-off of a tired one, which is an argument for maintained, quality vessels in small rooms.