Pick two goblets at random from the inventory record, not from the tank, and retrieve them with proper frost line technique. Everything the record says is in them should be there, correctly labelled, and nothing else. Then reverse the test: pick one physical goblet at random and confirm every straw or vial in it appears in the record. The reverse check matters because it catches the samples the inventory has forgotten, which the forward check never can. A spot check that fails in either direction means the whole inventory needs reconciling, and the reconciliation should happen vessel by vessel while samples remain where they are. The method sits inside our annual storage self-audit.